CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2013-06039
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Report Date
- March 4, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P930039
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS, AND ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT WHEN THE LEAD WAS REMOVED FROM THE PACKAGING, THE HELIX WAS ALREADY PARTIALLY EXTENDED. THE LEAD WAS ATTEMPTED BUT THEN REMOVED DUE TO HIGH IMPEDANCE AND DIFFICULTY EXTENDING THE HELIX. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.