FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153429 · Received June 8, 2013

Report

Report Number
2649622-2013-06017
Event Type
Injury
Date Received
June 8, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH. THE ANALYST COMMENTED THAT THE HELIX EXTENSION/RETRACTION AND LENGTH TESTING WERE PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 4296 IMPLANTABLE PACING LEAD (B)(6) 2013, 6945 IMPLANTABLE TACHY LEAD (B)(6) 1998. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS REMOVED FOR REPOSITIONING AND IT WAS THOUGHT THE TISSUE IMBEDDED AROUND THE HELIX WOULD INHIBIT THE RE-IMPLANT OF THE LEAD. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257079 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R (B)(4) IMPLANTABLE CARDIOVERTER DEFIBRILLATOR