FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3153415 · Received June 8, 2013

Report

Report Number
2182208-2013-01432
Event Type
Injury
Date Received
June 8, 2013
Report Date
June 25, 2013
Manufacturer
RICE CREEK MFG
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: IT WAS NOTED THAT THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING /OVERSTRESS, THE DISTAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS, THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO, THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS, AND THE VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: 7304 IMPLANTABLE PACEMAKER/CARDIO/DEFIB -UNKNOWN. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCE AND THERE WAS EXIT BLOCK. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258805 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX RICE CREEK MFG 419688

Patients

Seq Age Sex Outcome Treatment
1 00062 YR Hospitalization| R