FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3153404 · Received June 8, 2013

Report

Report Number
2182208-2013-01423
Event Type
Malfunction
Date Received
June 8, 2013
Report Date
March 20, 2013
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; PROGRAMMER DOES NOT POWER UP DUE TO A PRINTED CIRCUIT BOARD ASSEMBLY BEING OUT OF ELECTRICAL SPECIFICATION.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4): PRODUCT ID 2067 RF HEAD. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE PROGRAMMER DOES NOT POWER UP AND THERE IS NO DISPLAY. THE PROGRAMMER WAS RETURNED FOR REPAIR. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256396 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MEDTRONIC, INC. 2090AA

Patients

Seq Age Sex Outcome Treatment
1