FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3153343 · Received June 8, 2013

Report

Report Number
2182208-2013-01417
Event Type
Malfunction
Date Received
June 8, 2013
Report Date
March 20, 2013
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; RF (RADIO FREQUENCY) HEAD FAILED TWO UPLINK AMPLITUDE TESTS. RF HEAD CABLE FOUND OUT OF ELECTRICAL SPECIFICATION. ANALYSIS ALSO FOUND THE RUBBER PORTION OF THE RF HEAD LABEL IS MISSING.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE RF (RADIO FREQUENCY) HEAD MALFUNCTIONED FROM STERILE PROCESSING AND PACER CLINIC. NO ADDITIONAL INFORMATION WASOBTAINED THROUGH FOLLOW UP. THE RF HEAD WAS RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258096 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MEDTRONIC, INC. 2067

Patients

Seq Age Sex Outcome Treatment
1