FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153337 · Received June 8, 2013

Report

Report Number
2649622-2013-05962
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE DISTAL LEFT VENTRICULAR ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. THE ANALYST COMMENTED THAT THE HELIX WAS ABLE TO BE EXTENDED AND RETRACTED WITHIN SPECIFICATION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THERE WAS DIFFICULTY EXTENDING AND RETRACTING THE HELIX OF THE RIGHT VENTRICULAR LEAD. THE LEAD WAS OPENED BUT NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258094 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00073 YR