FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 3153334 · Received June 8, 2013

Report

Report Number
2649622-2013-05959
Event Type
Injury
Date Received
June 8, 2013
Report Date
April 4, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED. THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. VISUAL ANALYSIS NOTED SEVERE EXPLANT DAMAGE. IT WAS UNABLE TO BE DETERMINED WHERE THE ANCHOR SLEEVE WAS LOCATED AT TIME OF ANALYSIS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD AN APPARENT FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
258092 SPRINT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6945-65

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 1852 COMPETITOR IMPLANTABLE CARDIOVERTER DEFIBRILL