FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3153304 · Received June 8, 2013

Report

Report Number
2649622-2013-05918
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 15, 2013
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A TEAR. IT WAS ALSO NOTED THAT THE VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4) IMPLANTABLE PACING LEAD (B)(6) 2012. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD DISLODGED. AT THE LEAD REVISION THE PHYSICIAN HAD DIFFICULTY FREEING THE SUTURE SLEEVE. IT WAS THEN NOTED THAT THE LEAD HAD NOISE AND LOW IMPEDANCES. THE LEAD WAS THEN EXPLANTED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254987 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI45

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER