FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 3153303
·
Received June 8, 2013
Report
- Report Number
- 2182208-2013-01413
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Report Date
- March 20, 2013
- Manufacturer
- RICE CREEK MFG
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS FOUND INTERMITTENT TELEMETRY; RF (RADIO FREQUENCY) HEAD CABLE FOUND CUT. ANALYSIS ALSO FOUND THE LABEL WAS MISSING ITS COATING AND THE DEVICE HAD INTERNAL SIGNS OF WATER DAMAGE (CORROSION).
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Description of Event or Problem · 1
IT WAS REPORTED THE RF (RADIO FREQUENCY) HEAD MALFUNCTIONED FROM STERILE PROCESSING AND PACER CLINIC. NO ADDITIONAL INFORMATION WAS OBTAINED THROUGH FOLLOW UP. THE RF HEAD WAS RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256134 | CARELINK | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MFG | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |