FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3153268 · Received June 8, 2013

Report

Report Number
3004209178-2013-08898
Event Type
Injury
Date Received
June 8, 2013
Date of Event
April 7, 2013
Report Date
April 7, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS 5076 IMPLANTABLE PACING LEAD - (B)(6) 2013; 4296 IMPLANTABLE PACING LEAD - (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXHIBITED T-WAVE OVERSENSING (TWOS). THE DEVICE WAS EXPLANTED AND REPLACED, AND THE TWOS WAS FOUND TO HAVE RESOLVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXHIBITED T-WAVE OVERSENSING (TWOS). THE DEVICE WAS EXPLANTED AND REPLACED, AND THE TWOS WAS FOUND TO HAVE RESOLVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
255146 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D314TRM

Patients

Seq Age Sex Outcome Treatment
1 00063 YR Hospitalization| R 6935M IMPLANTABLE TACHY LEAD