FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3153226 · Received June 8, 2013

Report

Report Number
2649622-2013-05879
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 8, 2013
Report Date
March 12, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD IN SEVERAL PLACEMENTS HAD LOW SENSING AT 2.5 TO 5.0 MILLIVOLTS, INTERMITTENT CAPTURE AROUND 1.5 TO 2 VOLTS AT 0.5 MILLISECONDS AND IMPEDANCES APPROXIMATELY 1700 OHMS. DUE TO THE ELECTRICAL MEASUREMENTS THE PHYSICIAN ELECTED TO REMOVE THE RV LEAD AND IMPLANT A DIFFERENT LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254981 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407658

Patients

Seq Age Sex Outcome Treatment
1 00074 YR