CAPSURE SP NOVUS
Report
- Report Number
- 2649622-2013-05876
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 11, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD, THE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD WAS PLACED APICALLY BUT THE LEAD KICKED OUT TWICE FROM THE RV WHILE STILL POSITIONING. SO THE PHYSICIAN ELECTED TO REMOVE THE RV LEAD AND IMPLANT A DIFFERENT MODEL LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256029 | CAPSURE SP NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5092 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR |