FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3153191 · Received June 8, 2013

Report

Report Number
2649622-2013-05863
Event Type
Injury
Date Received
June 8, 2013
Date of Event
March 18, 2013
Report Date
March 19, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO AND THE PROXIMAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. A D284DRG, IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD).   (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSCHEDULED FOLLOW-UP THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DISPLAYED A LEAD INTEGRITY ALERT (LIA) DUE TO NOISE ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256039 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R