FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 3153188 · Received June 8, 2013

Report

Report Number
3004209178-2013-08892
Event Type
Injury
Date Received
June 8, 2013
Report Date
April 16, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. (B)(4): 4193 IMPLANTABLE PACING LEAD 2002-(B)(6), 5524M IMPLANTABLE PACING LEAD 2002-(B)(6). (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED AND REPLACED FOR NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI). THE DEVICE SUBSEQUENTLY WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THISEVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256038 CONCERTO CRT-D DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO C154DWK

Patients

Seq Age Sex Outcome Treatment
1 00049 YR Hospitalization| R 6945 IMPLANTABLE TACHY LEAD