SPRINT QUATTRO
Report
- Report Number
- 2649622-2013-05855
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- February 26, 2013
- Report Date
- March 8, 2013
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE ANALYST COMMENTED THAT HELIX EXTENSION/ RETRACTION AND LENGTH TESTING WERE PERFORMED AND ALL RESULTS, INCLUDING ELECTRICAL TESTING, WERE WITHIN SPECIFIED PARAMETERS.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD HELIX WOULD NOT EXTEND OR RETRACT. THE RV LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256075 | SPRINT QUATTRO | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00042 YR | 5592 IMPLANTABLE PACING LEAD |