ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-07890
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Report Date
- May 14, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION (B)(6) 2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2015 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE PUMP¿S VIBRATORY FEATURE WAS NOT FUNCTIONING. THE PUMP COVER WAS OPENED AND THE VIBRATION MOTOR PINS WERE FOUND TO BE MISALIGNED. THE PINS WERE REALIGNED, AND THE MOTOR FUNCTIONED PROPERLY DURING TESTING. UNRELATED TO THE COMPLAINT, THE DISPLAY SCREEN WAS DIM AND DISCOLORED. THE AUDIO BOLUS BUTTON COVER WAS TORN. THE BUTTON RESPONDED PROPERLY DURING TESTING. THE BATTERY COMPARTMENT WAS CRACKED. EVIDENCE OF MOISTURE CONTAMINATION WAS FOUND IN THE BATTERY COMPARTMENT. THE PUMP PASSED A LEAK TEST AND NO FURTHER MOISTURE INGRESS WAS FOUND IN THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A AUDIO TONE/VIBRATION (NO VIBRATION) ISSUE. THE REPORTER STATED THAT THE VIBRATION FEATURE ON THE PUMP WAS NOT WORKING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256045 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |