FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 3153122 · Received June 8, 2013

Report

Report Number
2649622-2013-05831
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 18, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT ID (B)(4) IMPLANTABLE PULSE GENERATOR CARDIO/DEFIB (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CARE ALERT TRIGGERED FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT GREATER THAN PROGRAMMED THRESHOLD. THEREFORE THE RV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CARE ALERT TRIGGERED FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT GREATER THAN PROGRAMMED THRESHOLD AND SHORT VENTRICLE-VENTRICLE (V-V) INTERVAL COUNTS. THEREFORE THE RV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256021 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R