SPRINT QUATTRO
Report
- Report Number
- 2649622-2013-05831
- Event Type
- Injury
- Date Received
- June 8, 2013
- Report Date
- March 18, 2013
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT ID (B)(4) IMPLANTABLE PULSE GENERATOR CARDIO/DEFIB (B)(6) 2010. (B)(4).
IT WAS REPORTED THAT THE CARE ALERT TRIGGERED FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT GREATER THAN PROGRAMMED THRESHOLD. THEREFORE THE RV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE CARE ALERT TRIGGERED FOR RIGHT VENTRICULAR (RV) LEAD PACING IMPEDANCE MEASUREMENT GREATER THAN PROGRAMMED THRESHOLD AND SHORT VENTRICLE-VENTRICLE (V-V) INTERVAL COUNTS. THEREFORE THE RV LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256021 | SPRINT QUATTRO | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694765 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |