FDA Adverse Event Injury Summary report: N

PROTECTA XT DR

MDR report key: 3153075 · Received June 8, 2013

Report

Report Number
3002807576-2013-00012
Event Type
Injury
Date Received
June 8, 2013
Date of Event
March 5, 2013
Report Date
March 7, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD. ANALYSIS OF THE DEVICE MEMORY INDICATED THE CRITERIA FOR THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT WERE MET. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO EMI INTERFERENCE. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256163 PROTECTA XT DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND D354DRG

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R