PROTECTA XT DR
Report
- Report Number
- 3002807576-2013-00012
- Event Type
- Injury
- Date Received
- June 8, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 7, 2013
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD. ANALYSIS OF THE DEVICE MEMORY INDICATED THE CRITERIA FOR THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT WERE MET. (B)(4).
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO EMI INTERFERENCE. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 256163 | PROTECTA XT DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | IPG MFG SWITZERLAND | D354DRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | Hospitalization| R |