FDA Adverse Event Malfunction Summary report: N

RELIA SR

MDR report key: 3153037 · Received June 8, 2013

Report

Report Number
3008973940-2013-00064
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 14, 2013
Report Date
March 14, 2013
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
NVZ
PMA / PMN Number
P980035
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE SET SCREW WAS DAMAGED AND THE SET SCREW WAS FOUND IN THE CONNECTOR BORE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT THE SETSCREW DID NOT WORK SO IT WAS NOT POSSIBLE TO SCREW THE LEAD INTO THE HEADER. THE DEVICE WAS NOT USED AND ANOTHER DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
256149 RELIA SR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC SINGAPORE OPERATIONS RESR01

Patients

Seq Age Sex Outcome Treatment
1