FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3153023 · Received June 8, 2013

Report

Report Number
2649622-2013-05783
Event Type
Injury
Date Received
June 8, 2013
Date of Event
April 26, 2013
Report Date
April 29, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND, THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO FLATTENING. NO INSULATION BREACH WAS NOTED; HOWEVER THE PROXIMAL AND DISTAL CONDUCTORS WERE PINCHED/FLATTENED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD SHOWED INSULATION DAMAGE DURING AN X-RAY. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254473 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419678

Patients

Seq Age Sex Outcome Treatment
1 00062 YR Hospitalization| R