FDA Adverse Event
Injury
Summary report: N
ATTAIN ABILITY
MDR report key: 3153023
·
Received June 8, 2013
Report
- Report Number
- 2649622-2013-05783
- Event Type
- Injury
- Date Received
- June 8, 2013
- Date of Event
- April 26, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND, THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO FLATTENING. NO INSULATION BREACH WAS NOTED; HOWEVER THE PROXIMAL AND DISTAL CONDUCTORS WERE PINCHED/FLATTENED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD SHOWED INSULATION DAMAGE DURING AN X-RAY. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254473 | ATTAIN ABILITY | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 419678 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00062 YR | Hospitalization| R |