PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2182208-2013-01366
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 11, 2013
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
FURTHER ANALYSIS WAS PERFORMED ON THE MAIN PRINTED CIRCUIT BOARD ASSEMBLY (PCBA). THIS ANALYSIS COULD NOT CONFIRM THE REPORTED FAILURE OF POWER UP IN CAL MODE, HOWEVER IT WAS CONFIRMED THAT THE DEVICE WAS SIGNIFICANTLY OUT OF CALIBRATION. RE-CALIBRATION PRODUCED DEVICE OPERATION WITHIN SPECIFICATION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; THE DEVICE POWERED UP IN "CAL" MODE. THE MAIN PRINTED CIRCUIT BOARD ASSEMBLY WAS FOUND OUT OF ELECTRICAL SPECIFICATION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED WHEN THE ON BUTTON IS PRESSED THE EPG (EXTERNAL PULSE) PACEMAKER DOES NOT START PACING, IT GIVES A "CAL" ERROR IN THE RAP (RAPID ATRIAL PACING) WINDOW. ONCE THE ENABLE/DISABLE BUTTON IS PRESSED FOR THE RAP, THE MESSAGE CHANGES TO "111". THE EPG WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254866 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC, INC. | 5348 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |