FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 3152903
·
Received June 8, 2013
Report
- Report Number
- 2182208-2013-01364
- Event Type
- Malfunction
- Date Received
- June 8, 2013
- Date of Event
- March 15, 2013
- Report Date
- March 15, 2013
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, AS A RESULT THE OVERLAY/BEZEL ASSEMBLY WAS REPLACED AND THE STYLUS WAS CALIBRATED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE TOUCH PEN COULD NOT BE CALIBRATED, AND THE TOUCH PEN MUST BE TWO TO THREE INCHES AWAY FROM THE SCREEN IN ORDER TO ACTIVATE THE TASK ICON. IN ADDITION, SOME AREAS OF THE SCREEN DID NOT GET A RESPONSE. THE PROGRAMMER WAS RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 254735 | CARELINK | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | MEDTRONIC, INC. | 2090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 2067 PROGRAMMER RF (RADIO-FREQUENCY) HEAD |