FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 3152814 · Received June 8, 2013

Report

Report Number
3004209178-2013-08847
Event Type
Injury
Date Received
June 8, 2013
Report Date
March 14, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. CONCOMITANT PRODUCT: 6947, IMPLANTABLE TACHY LEAD, (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS REMOVED AND REPLACED DUE TO SUSPECTED PREMATURE BATTERY DEPLETION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257478 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00071 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD