FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3152755 · Received June 8, 2013

Report

Report Number
2649622-2013-05686
Event Type
Malfunction
Date Received
June 8, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVALUATION SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. IT WAS ALSO NOTED THAT DISTAL ELECTRODE WAS COVERED WITH BLOOD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT THE RIGHT VENTRICULAR (RV) LEAD WAS EXHIBITING NOISE. THE LEAD WAS REPOSITIONED BUT THE NOISE CONTINUED. THE LEAD WAS REMOVED AND REPLACED WITH A NEW RV LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
257472 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00079 YR