ATTAIN ABILITY
Report
- Report Number
- 2649622-2013-06718
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 5, 2013
- Report Date
- April 8, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4).
PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR WAS EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR OF THE LEAD WAS OBSTRUCTED DUE TO A STYLET/GUIDEWIRE STUCK IN THE LUMEN. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE GUIDEWIRE GOT STUCK IN THE LEAD AND COULD NOT BE REMOVED. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253191 | ATTAIN ABILITY | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 419688 |