FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 3152648 · Received June 7, 2013

Report

Report Number
2649622-2013-06718
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 5, 2013
Report Date
April 8, 2013
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THAT THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR WAS EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR OF THE LEAD WAS OBSTRUCTED DUE TO A STYLET/GUIDEWIRE STUCK IN THE LUMEN. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT, THE GUIDEWIRE GOT STUCK IN THE LEAD AND COULD NOT BE REMOVED. THE LEAD WAS NOT USED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253191 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00088 YR