Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE PROXIMAL AND DISTAL CONDUCTORS OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. THERE WAS ALSO BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE DISTAL LV (LEFT VENTRICULAR) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH. THE INNER AND OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A TEAR. THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE OUTER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING. ANALYST COMMENTED THE LEAD WAS RETURNED WITH SEVERE EXPLANT DAMAGE. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 694765, IMPLANTABLE TACHY LEAD, IMPLANTED: (B)(6) 2004; D224DRG, IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), IMPLANTED: (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS POOR SENSING AND HIGH THRESHOLD ON THE RIGHT ATRIAL LEAD. DURING THE EXTRACTION OF THE ATRIAL LEAD, THE RIGHT VENTRICULAR (RV) LEAD INSULATION AND SUPERIOR VENA CAVA (SVC) COIL WERE COMPROMISED. BOTH LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252118 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-45

Patients

Seq Age Sex Outcome Treatment
1 00054 YR Hospitalization| R