FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 3152541 · Received June 7, 2013

Report

Report Number
2649622-2013-06769
Event Type
Injury
Date Received
June 7, 2013
Report Date
April 9, 2013
Manufacturer
MPRI
Product Code
LWP
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 7299, IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), IMPLANTED: (B)(6) 2008; 5076, IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS EXPERIENCING HIGH IMPEDANCE AND HAD DISLODGED. THE LEAD WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252079 ATTAIN OTW PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MPRI 419388

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD