FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3152453 · Received June 7, 2013

Report

Report Number
2649622-2013-06793
Event Type
Injury
Date Received
June 7, 2013
Report Date
April 26, 2013
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED LOWER RANGE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 4076545 IMPLANTABLE PACING LEAD, (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS A POLARITY SWITCH ALONG WITH LOW AND HIGH IMPEDANCE READINGS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252812 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 407652

Patients

Seq Age Sex Outcome Treatment
1 00080 YR ADDR01 IMPLANTABLE PULSE GENERATOR