FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3152428 · Received June 7, 2013

Report

Report Number
2183613-2013-00673
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 19, 2013
Report Date
April 19, 2013
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS COULD NOT CONFIRM THE CUSTOMER COMMENT THAT THE GENERATOR'S SENSE AND PACE PORTS WERE LOOSE. ANALYSIS DID FIND THAT THE UPPER AND LOWER CASES WERE BROKEN, ONE BAIL COVER WAS BROKEN AND ONE MISSING AND THAT THE KEYBOARD WAS SCRATCHED. (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S SENSE AND PACE PORTS WERE LOOSE, AND WERE INTERMITTENTLY "FLYING." THE GENERATOR WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR'S SENSE AND PACE PORTS WERE LOOSE, AND WERE INTERMITTENTLY "FLYING." THE GENERATOR WAS RETURNED FOR SERVICE. IT WAS INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252722 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1