SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2013-06819
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- March 22, 2013
- Report Date
- March 22, 2013
- Manufacturer
- MPRI
- Product Code
- NVY
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX OF THE LEAD WAS EXTRINSICALLY BENT. THE HELIX OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE RIGHT VENTRICULAR (RV) LEAD WAS IMPLANTED AND SHOT OUT ABRUPTLY. THERE WAS LOW IMPEDANCE, HIGH THRESHOLD AND POOR SENSING MEASURED. THE LEAD WAS PLACED IN TWO OTHER LOCATIONS WITH SIMILAR RESULTS. THE LEAD WAS ATTEMPTED AND NOT USED. ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252715 | SPRINT QUATTRO SECURE | DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES | NVY | MPRI | 6947M55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00051 YR |