PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2013-00680
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 24, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Removal / Correction Number
- Z-0430-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE MAIN PRINTED CIRCUIT BOARD WAS OUT OF SPECIFICATION. IT WAS ALSO NOTED THAT THE RING AND SIDE BAIL COVERS WERE CONTAMINATED, THE BATTERY CONTACTS WERE COMPRESSED, THE RING BAIL WAS BENT, THE BATTERY DRAWER WAS CONTAMINATED, THE KEYBOARD WAS SCRATCHED AND THE DISPLAY WAS MISSING PIXELS.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR COULD NOT PASS ITS INITIAL SELF-TEST. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.