FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3152342 · Received June 7, 2013

Report

Report Number
2183613-2013-00680
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 24, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE MAIN PRINTED CIRCUIT BOARD WAS OUT OF SPECIFICATION. IT WAS ALSO NOTED THAT THE RING AND SIDE BAIL COVERS WERE CONTAMINATED, THE BATTERY CONTACTS WERE COMPRESSED, THE RING BAIL WAS BENT, THE BATTERY DRAWER WAS CONTAMINATED, THE KEYBOARD WAS SCRATCHED AND THE DISPLAY WAS MISSING PIXELS.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR COULD NOT PASS ITS INITIAL SELF-TEST. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.