FDA Adverse Event Malfunction Summary report: N

REVEAL XT

MDR report key: 3152305 · Received June 7, 2013

Report

Report Number
3004209178-2013-09056
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 11, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
DSI
PMA / PMN Number
K103764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED UNDERSENSING INFORMATION. THE ANALYST COMMENTED REGARDING FALSE ASYSTOLE DUE TO LOSS OF CONTACT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR WAS EXPERIENCING UNDERSENSING AND "LOSS OF CONTACT." THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252833 REVEAL XT DETECTOR AND ALARM, ARRHYTHMIA DSI MEDTRONIC MED REL MEDTRONIC PUERTO RICO 9529

Patients

Seq Age Sex Outcome Treatment
1