FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 3152291 · Received June 7, 2013

Report

Report Number
3004209178-2013-09060
Event Type
Injury
Date Received
June 7, 2013
Report Date
March 5, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE FOUND A HIGH CURRENT DRAIN CONDITION DUE TO CURRENT LEAKAGE IN A CERAMIC CAPACITOR.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED EARLY BATTERY DEPLETION. THEREFORE, THE DEVICE WAS REMOVED AND REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252951 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154VWC

Patients

Seq Age Sex Outcome Treatment
1 00048 YR Hospitalization| R 6947 IMPLANTABLE TACHY LEAD