FDA Adverse Event Malfunction Summary report: N

DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE

MDR report key: 3152284 · Received June 7, 2013

Report

Report Number
3008973940-2013-00087
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 27, 2013
Report Date
April 27, 2013
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND THE INNER TUBING OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A TEAR, THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS, THE INNER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO PULLING/STRETCHING/OVERSTRESS, AND THE INNER TUBING OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE INNER INSULATION OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER PLACING THE RIGHT VENTRICULAR (RV) LEAD, MEASUREMENT VIA THE PACING SYSTEM ANALYZER (PSA) WAS PERFORMED AND THERE WAS NOTED LOW R-WAVES AND THERE WAS NO CAPTURE AT 5V. THEREFORE THE RV LEAD WAS REPOSITIONED HOWEVER, EXTENSION OF THE HELIX COULD NOT BE CONFIRMED FOLLOWING ROTATION WITH THE PINCH ON TOOL. THE RV LEAD WAS REMOVED AND HELIX EXTENSION/RETRACTION TEST WAS PERFORMED AND THE HELIX WOULD NOT EXTEND AFTER SEVERAL ROTATIONS. THE RV LEAD WAS NOT USED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252661 DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MEDTRONIC SINGAPORE OPERATIONS 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00073 YR