CARELINK
Report
- Report Number
- 2182208-2013-01573
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 24, 2013
- Report Date
- April 24, 2013
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING, THE HEAD CABLE WAS FOUND TO BE TWISTED AND THE HEAD FAILED THREE UPLINK AMPLITUDE TESTS. ADDITIONALLY, THE RUBBER PORTION OF THE HEAD LABEL WAS FOUND TO BE COMING LOOSE AND IT, AS WELL AS THE HEAD'S CABLE WAS REPLACED. (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253048 | CARELINK | PROGRAMMER, PACEMAKER | KRG | MEDTRONIC, INC. | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |