FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 3152275
·
Received June 7, 2013
Report
- Report Number
- 2182208-2013-01571
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Report Date
- March 21, 2013
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE DEVICE FAILED TELEMETRY TESTING DUE TO THE CABLE BEING OUT OF ELECTRICAL SPECIFICATION. CONCOMITANT PRODUCTS: PRODUCT ID 209031 PROGRAMMER; PRODUCT ID 229047 ANALYZER. (B)(4).
Additional Manufacturer Narrative · 1
FURTHER REVIEW PROMPTED A CHANGE IN THE DEVICE ANALYSIS RESULTS.
Description of Event or Problem · 1
THE PROGRAMMER AND RADIOFREQUENCY (RF) HEAD WERE RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.