FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 3152252 · Received June 7, 2013

Report

Report Number
3004209178-2013-08780
Event Type
Injury
Date Received
June 7, 2013
Report Date
April 4, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR A WARRANTY CREDIT. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252871 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Hospitalization| R