FDA Adverse Event Injury Summary report: N

INSYNC III PROTECT

MDR report key: 3152112 · Received June 7, 2013

Report

Report Number
9614453-2013-01078
Event Type
Injury
Date Received
June 7, 2013
Report Date
February 28, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR DUE TO POSSIBLE PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. DURING THE REPLACEMENT PROCEDURE, THE PHYSICIAN FOUND THAT THE ATRIAL LEAD THRESHOLD HAD INCREASED. A NEW LEAD WAS ATTEMPTED TO BE IMPLANTED, BUT FAILED. THAT LEAD WAS NOT USED AND THE EXISTING LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252802 INSYNC III PROTECT DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND 7285

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 4574 IMPLANTABLE LEAD