ATTAIN ABILITY PLUS
Report
- Report Number
- 2649622-2013-05546
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 1, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD ON THE PROXIMAL AND DISTAL CONDUCTORS OF THE LEAD AND IT WAS NOT OBSTRUCTED.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT DURING THE PLACEMENT OF THE IMPLANTABLE PACING LEAD THE IMPEDANCE WAS ALWAYS ABOVE 3000 OHMS. THE PHYSICIAN DECIDED TO USE ANOTHER LEAD, WHICH WAS PLACED WITHOUT INCIDENT, AND THE LEAD EXHIBITING THE HIGH IMPEDANCE WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252879 | ATTAIN ABILITY PLUS | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 429688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|