FDA Adverse Event Injury Summary report: N

ENRHYTHM

MDR report key: 3152036 · Received June 7, 2013

Report

Report Number
3004209178-2013-08795
Event Type
Injury
Date Received
June 7, 2013
Date of Event
March 8, 2013
Report Date
March 8, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 1888TC COMPETITOR IMPLANTABLE PACING LEAD (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) HAD A POWER ON RESET (POR). IT WAS NOTED THAT THE PATIENT WAS UNDERGOINGRADIATION THERAPY. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252878 ENRHYTHM PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO P1501DR

Patients

Seq Age Sex Outcome Treatment
1 00071 YR Required Intervention 5076 IMPLANTABLE PACING LEAD