FDA Adverse Event Injury Summary report: N

GEM III

MDR report key: 3152023 · Received June 7, 2013

Report

Report Number
9614453-2013-01083
Event Type
Injury
Date Received
June 7, 2013
Report Date
March 21, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. THE TIME OF ELECTIVE REPLACEMENT INDICATOR (ERI) IN SAVE TO DISK WAS (B)(6) 2013, WITH A VOLTAGE OF LESS THAN OR EQUAL TO 2.55 VOLTS ON THE DEVICE. THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)  REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLIER THAN EXPECTED. THE ICD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253053 GEM III DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND 7231CX

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R 6947 IMPLANTABLE TACHY LEAD