FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3152018 · Received June 7, 2013

Report

Report Number
2183613-2013-00533
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
March 25, 2013
Report Date
March 25, 2013
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; LCD (LIQUID CRYSTAL DISPLAY) IS OUT OF ELECTRICAL SPECIFICATION (MISSING PIXELS).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING INSPECTION OF THE EXTERNAL PULSE GENERATOR IT WAS DISCOVERED THAT THERE WERE MISSING SEGMENTS IN THE LOWER DISPLAY. THE GENERATOR WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.