FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3151983 · Received June 7, 2013

Report

Report Number
2183613-2013-00535
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
March 12, 2013
Report Date
March 12, 2013
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT, THE BATTERY CONTACTS WERE OUT OF SPECIFICATION. IT WAS ALSO NOTED THAT THE LOWER CASE WAS BROKEN, THE BATTERY DRAWER WAS BROKEN, AND THE SIDE BAIL COVERS WERE CONTAMINATED. (B)(4).

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WILL NOT POWER UP. IT WAS NOTED THE METAL BATTERY CONTACT SEEMS BROKEN. THE DEVICE WAS RETURNED FOR REPAIR. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252436 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1