DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Report
- Report Number
- 2649622-2013-05580
- Event Type
- Injury
- Date Received
- June 7, 2013
- Date of Event
- January 1, 2013
- Report Date
- March 1, 2013
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
THIS INFORMATION IS BASED ENTIRELY ON JOURNAL LITERATURE. THIS EVENT OCCURRED OUTSIDE THE U.S. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS LIMITED DUE TO CONFIDENTIALITY CONCERNS. REFERENCED ARTICLE: OCCURRENCE OF PHRENIC NERVE STIMULATION IN CARDIAC RESYNCHRONIZATION THERAPY PATIENTS: THE ROLE OF LEFT VENTRICULAR LEAD TYPE AND PLACEMENT SITE. EUROPACE: EUROPEAN PACING, ARRHYTHMIAS, AND CARDIAC ELECTROPHYSIOLOGY: JOURNAL OF THE WORKING GROUPS ON CARDIAC PACING, ARRHYTHMIAS, AND CARDIAC CELLULAR ELECTROPHYSIOLOGY OF THE EUROPEAN SOCIETY OF CARDIOLOGY. 2013;15(1):77-82. (B)(4).
A JOURNAL ARTICLE WAS REVIEWED WHICH CONTAINED INFORMATION REGARDING THIS LEAD. THE PATIENT EXPERIENCED DIAPHRAGMATIC MUSCLE STIMULATION ONE MONTH POST IMPLANT. THE LEFT VENTRICULAR (LV) OUTPUT WAS ADJUSTED AND THE LEAD REMAINED IN USE. UPON ANOTHER OCCASION OF STIMULATION, THE POLARITY OF THE LEFT VENTRICULAR (LV) WAS CHANGED TO RESOLVE THE ISSUE AND THE LEAD REMAINED IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252650 | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 439688C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00067 YR | Hospitalization| R |