FDA Adverse Event
Malfunction
Summary report: N
PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
MDR report key: 3151919
·
Received June 7, 2013
Report
- Report Number
- 2183613-2013-00544
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 15, 2013
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT; VENTRICLE OUTPUT CONNECTOR IS BROKEN. ANALYSIS ALSO FOUND THE UPPER AND LOWER CASES ARE BROKEN, BOTH BAIL COVERS ARE BROKEN AND CONTAMINATED, RING COVER IS CONTAMINATED, BATTERY CONTACTS ARE COMPRESSED, AND THE KEYBOARD IS SCRATCHED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR HAD A BROKEN CONNECTOR. THE GENERATOR WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252342 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5388 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |