FDA Adverse Event Injury Summary report: N

CONCERTO II CRT-D

MDR report key: 3151857 · Received June 7, 2013

Report

Report Number
3004209178-2013-08806
Event Type
Injury
Date Received
June 7, 2013
Date of Event
January 31, 2013
Report Date
March 27, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0110-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 6944 IMPLANTABLE DEFIB LEAD (B)(6) 2003. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY (B)(4) - THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED NORMAL BATTERY DEPLETION. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.  (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE EXPERIENCED EARLY ELECTIVE INDICATOR (ERI) DUE TO HIGH OUTPUT ON THE LEFT VENTRICULAR LEAD. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252627 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D274TRK

Patients

Seq Age Sex Outcome Treatment
1 00081 YR Hospitalization| R 4193-78 IMPLANTABLE PACING LEAD