FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3151852 · Received June 7, 2013

Report

Report Number
2182208-2013-01232
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
April 15, 2013
Report Date
April 15, 2013
Manufacturer
RICE CREEK MANUFACTURING
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE RADIO FREQUENCY (RF) PROGRAMMER HEAD WAS RETURNED AND ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT IT DID NOT WORK. IT WAS UNABLE TO INTERROGATE AND ITS UPLINK FUNCTIONAL TESTS WERE OUT OF SPECIFICATION. THE PROGRAMMER HEAD'S CABLE WAS REPLACED AND THE HEAD THEN PASSED ALL FUNCTIONAL AND SYSTEMS TESTS. (B)(4).

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: PRODUCT ID 2090AM PROGRAMMER; PRODUCT ID 229047 SOFTWARE ANALYZER. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE PROGRAMMER HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTEDAS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROGRAMMER'S PRINTER WAS NOT WORKING. IT WAS FURTHER REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE PROGRAMMER AND THE PROGRAMMER HEAD WERE RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTEDAS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253254 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MANUFACTURING 2067

Patients

Seq Age Sex Outcome Treatment
1