CARELINK
Report
- Report Number
- 2182208-2013-01232
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- RICE CREEK MANUFACTURING
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT EVENT SUMMARY: THE RADIO FREQUENCY (RF) PROGRAMMER HEAD WAS RETURNED AND ANALYSIS CONFIRMED THE CUSTOMER COMMENT THAT IT DID NOT WORK. IT WAS UNABLE TO INTERROGATE AND ITS UPLINK FUNCTIONAL TESTS WERE OUT OF SPECIFICATION. THE PROGRAMMER HEAD'S CABLE WAS REPLACED AND THE HEAD THEN PASSED ALL FUNCTIONAL AND SYSTEMS TESTS. (B)(4).
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: PRODUCT ID 2090AM PROGRAMMER; PRODUCT ID 229047 SOFTWARE ANALYZER. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE PROGRAMMER HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTEDAS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE PROGRAMMER'S PRINTER WAS NOT WORKING. IT WAS FURTHER REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE PROGRAMMER AND THE PROGRAMMER HEAD WERE RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTEDAS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253254 | CARELINK | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MANUFACTURING | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |