CARELINK
Report
- Report Number
- 2182208-2013-01237
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 15, 2013
- Report Date
- April 15, 2013
- Manufacturer
- RICE CREEK MFG
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE LABEL BACKING WAS MISSING, THE CABLE WAS PULLING APART NEAR THE BASE OF THE DEVICE, THE DEVICE TESTED OUT OF SPECIFICATION: ELECTROMAGNETIC FIELD IMMUNITY AND UPLINK TESTS, AND THE LENS WAS CRACKED ON THE UPPER CASE.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCTS: PRODUCT ID 2090AM PROGRAMMER. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RADIO FREQUENCY PROGRAMMER HEAD WAS NOT WORKING. THE PROGRAMMER HEAD WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.