FDA Adverse Event Malfunction Summary report: N

SENSIA

MDR report key: 3151834 · Received June 7, 2013

Report

Report Number
9614453-2013-01085
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE DEVICE WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THE RETURNED DEVICE INDICATED A DAMAGED GROMMET AND FOREIGN MATERIAL ON SET SCREW.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO IMPLANT THE SCREW THAT IS USED TO HOLD THE LEAD IN PLACE WAS LOOSE IN THE SEALED BOX. THE DEVICE WAS NOT USED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252746 SENSIA PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND SEDR01

Patients

Seq Age Sex Outcome Treatment
1