FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 3151769 · Received June 7, 2013

Report

Report Number
3004209178-2013-08812
Event Type
Injury
Date Received
June 7, 2013
Report Date
March 1, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. CONCOMITANT PRODUCT: 5076, IMPLANTABLE PACING LEAD, IMPLANTED (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC DEFIBRILLATOR (ICD) DID NOT MEET EXPECTED LONGEVITY. EARLY BATTERY DEPLETION IS ALLEGED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253220 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Hospitalization| R 6947, IMPLANTABLE DEFIBRILLATION LEAD